ZHCU623 January 2019
In total, 10 patches were tested for accuracy within the liquid oil-bath. The results are given in Figure 25. The patches were assessed to be well within the specified ASTM requirements for medical-grade accuracy given in Table 2, therefore no additional calibration was applied to the TMP117's offset register. For products actually intended for use in a medical setting, packaging must be included during accuracy testing to ensure total system accuracy.