Measure the voltage from 1P to 1N (VSS) and enter this value in the Applied Cell 1 Voltage field and select the Calibrate Voltage box.
Measure the voltage from 6P to 1N (VSS) and enter this value in the Applied Battery Voltage field and select the Calibrate Battery Voltage box.
Measure the voltage from Pack+ to Pack– and enter this value in the Applied Pack Voltage field and select the Calibrate Pack Voltage box. If the voltage is not present, then turn the charge and discharge FETs on by entering a 0x0022 command in the Manufacturer Access register on the Register screen.
If the device is configured for 7S, measure the voltage from 7P to 1N (VSS) and enter this value in the Applied Cell-7 Positive Terminal Voltage field and select the Calibrate Battery Voltage box.
Press the Calibrate Gas Gauge button to calibrate the voltage measurement system.
Deselect the Calibrate Voltage boxes after voltage calibration has completed.