Purpose: Adjusts the device's internal cap
array setting for tuning the HF crystal clock accuracy to meet the crystal clock
accuracy spec.
Pass Condition: Commanded RF output
frequency is within crystal data sheet's specified tolerances with final cap array
setting.
Procedure:
- Step 1: Connect the instruments
as shown in Figure 6-1 and Figure 6-2.
- Step 2: Open SmartRF Studio and
select the desired frequency on RF Parameters.
- Step 3: Checkmark (enable) the
Cap-Array Tuning. Start with the Cap-Array at 0.
- Step 4: At the Continous TX tab,
make sure the TX is unmodulated and click start.
- Step 5: Measure the output frequency on the spectrum analyzer.
- Step 6: If the measured frequency
is out of tolerance for the frequency commanded, stop the output signal and then
change the Cap-Array (Delta) value depending on required adjustment. If the
measured frequency is above the commanded frequency, increase the Cap-Array
Setting. If measured frequency is below the commanded frequency, decrease the
Cap-Array Setting.
- Step 7: Enable the RF output and
remeasure the output frequency.
- Step 8: Repeat Steps 6-7 until a
Cap-Array Delta value which generates a measured frequency that is as close to
the commanded frequency as possible is found.
Note: Ensure the RBW and frequency span of the
spectrum analyzer are reduced to ensure the measured frequency resolution is at
least within the hundreds of Hz. Also, it is recommended to use a spectrum analyzer
with a precision frequency reference for best accuracy. For custom applications,
please refer to
SWRA495 for help with crystal selection and implementation.
Table 6-2 HF Oscillator Results
|
Cap-Array Value |
Chosen Frequency |
Output Frequency |
Frequency Delta |
PPM |
| 1 |
|
|
|
|
|
| 2 |
|
|
|
|
|
| 3 |
|
|
|
|
|
| 4 |
|
|
|
|
|