AFE159RP4

正在供货

具有呼吸、起搏器、导联脱落和导联质量检测以及 FIFO 功能的低功耗 4 通道 ECG AFE

产品详情

Applications AEDs, Bedside patient monitoring and diagnostic ECG, ECG patch, EEG, EMG, Event, Evoked audio potential (EAP), Holter monitor and multi-lead patch, Portable telemetry, Respiration, Sleep study monitor, Telemedicine bispectral index (BIS) Number of input channels 4 Resolution (Bits) 24 Interface type SPI Operating temperature range (°C) -20 to 85 Rating Catalog
Applications AEDs, Bedside patient monitoring and diagnostic ECG, ECG patch, EEG, EMG, Event, Evoked audio potential (EAP), Holter monitor and multi-lead patch, Portable telemetry, Respiration, Sleep study monitor, Telemedicine bispectral index (BIS) Number of input channels 4 Resolution (Bits) 24 Interface type SPI Operating temperature range (°C) -20 to 85 Rating Catalog
VQFN (RGZ) 48 49 mm² (7 mm × 7 mm) DSBGA (YAF) 49 13.387056 mm² (3.694 mm × 3.624 mm)
  • Integrated Signal Chain for ECG, Pace Detection, and Respiration Measurement
  • ECG Receiver
    • Four high resolution channels at low power of 0.42mW/channel
    • Flexible four leads selectable from six electrodes
    • Programmable gain: 1.25 to 9
    • Input-referred noise: 4µVPP in 150Hz BW
    • Differential input range: ±1V with Gain = 4
    • CMRR: -140dB
    • Data rate: 125SPS to 128kSPS
  • Pace Detection
    • On-chip digital pace detection algorithm on programmable two leads
    • High-speed 128kSPS pace output on two channels for software pace detection
  • Respiration
    • Low-noise of 24mΩPP with 2kΩ body impedance and 1kΩ defibrillator protection resistor on each electrode
    • Supports Sine and Square wave excitation
  • Other Features
    • Built-in right leg drive amplifier steerable to any electrode
    • DC lead-off detection, AC lead impedance detection, Wilson Center Terminal (WCT), Goldberger Central Terminals (GCT), test signals
    • Battery voltage monitoring
    • Flexible power-down and standby modes
    • Built-in oscillator, PLL, and reference
    • 1k sample main FIFO and 2k sample pace FIFO
    • SPI-compatible serial interface
    • Analog supply voltage 1: 3.15V to 5.25V
    • Analog supply voltage 2: 1.7V to 1.9V
    • I/O supply voltage: 1.65V to 3.6V
  • Supports systems meeting AAMI EC11, AMI EC13, AMI EC38, IEC60601-1, IEC60601-2-25, IEC60601-2-27, and IEC60601-2-51 standards
  • Integrated Signal Chain for ECG, Pace Detection, and Respiration Measurement
  • ECG Receiver
    • Four high resolution channels at low power of 0.42mW/channel
    • Flexible four leads selectable from six electrodes
    • Programmable gain: 1.25 to 9
    • Input-referred noise: 4µVPP in 150Hz BW
    • Differential input range: ±1V with Gain = 4
    • CMRR: -140dB
    • Data rate: 125SPS to 128kSPS
  • Pace Detection
    • On-chip digital pace detection algorithm on programmable two leads
    • High-speed 128kSPS pace output on two channels for software pace detection
  • Respiration
    • Low-noise of 24mΩPP with 2kΩ body impedance and 1kΩ defibrillator protection resistor on each electrode
    • Supports Sine and Square wave excitation
  • Other Features
    • Built-in right leg drive amplifier steerable to any electrode
    • DC lead-off detection, AC lead impedance detection, Wilson Center Terminal (WCT), Goldberger Central Terminals (GCT), test signals
    • Battery voltage monitoring
    • Flexible power-down and standby modes
    • Built-in oscillator, PLL, and reference
    • 1k sample main FIFO and 2k sample pace FIFO
    • SPI-compatible serial interface
    • Analog supply voltage 1: 3.15V to 5.25V
    • Analog supply voltage 2: 1.7V to 1.9V
    • I/O supply voltage: 1.65V to 3.6V
  • Supports systems meeting AAMI EC11, AMI EC13, AMI EC38, IEC60601-1, IEC60601-2-25, IEC60601-2-27, and IEC60601-2-51 standards

The AFE1594 is a family of multichannel, simultaneous sampling, 24-bit, delta-sigma (ΔΣ) analog-to-digital converters (ADCs) with built-in programmable gain Instrumentation Amplifiers (INAs), internal reference, and an on-chip PLL. The AFE supports digital pace pulse detection, thoracic impedance measurement and incorporates all of the features that are commonly required in medical electrocardiogram (ECG) and electroencephalogram (EEG) applications. Multiple AFE159x devices can be cascaded in high channel count systems. With high levels of integration and exceptional performance, the AFE159x enables the development of scalable medical instrumentation systems at significantly reduced size, power, and overall cost.

The AFE1594 is a family of multichannel, simultaneous sampling, 24-bit, delta-sigma (ΔΣ) analog-to-digital converters (ADCs) with built-in programmable gain Instrumentation Amplifiers (INAs), internal reference, and an on-chip PLL. The AFE supports digital pace pulse detection, thoracic impedance measurement and incorporates all of the features that are commonly required in medical electrocardiogram (ECG) and electroencephalogram (EEG) applications. Multiple AFE159x devices can be cascaded in high channel count systems. With high levels of integration and exceptional performance, the AFE159x enables the development of scalable medical instrumentation systems at significantly reduced size, power, and overall cost.

下载 观看带字幕的视频 视频
申请了解更多信息

可提供完整的数据表。立即申请

您可能感兴趣的相似产品

open-in-new 比较替代产品
功能与比较器件相同且具有相同引脚。
AFE159P4 正在供货 具有起搏器、导联脱落和导联质量检测以及 FIFO 功能的低功耗 4 通道 ECG AFE Same device minus respiration.
AFE1594 正在供货 具有导联脱落和导联质量检测以及 FIFO 功能的低功耗 4 通道 ECG AFE Same device minus respiration and pace detection.

技术文档

未找到结果。请清除搜索并重试。
查看全部 1
顶层文档 类型 标题 格式选项 下载最新的英语版本 日期
* 数据表 AFE159x Low-Power, 4-Channel, 24-Bit Analog Front-Ends for Bio-Potential Measurements 数据表 (Rev. A) PDF | HTML 2024-11-12

设计与开发

如需其他信息或资源,请点击以下任一标题进入详情页面查看(如有)。

评估板

AFE1594EVM — AFE1594 评估模块

AFE1594 评估套件是一款开发平台,可用于评估适用于患者监护和诊断应用的 AFE1594 的性能。AFE1594 是多通道同步采样 24 位 Δ-Σ 模数转换器 (ADC) 系列,此系列 ADC 具有内置可编程增益仪表放大器 (INA),内部基准以及片上时钟。AFE 将心电图 (ECG) 和脑电图 (EEG) 等医疗应用所需的所有常见功能集于一身。凭借高集成度以及出色的性能,AFE1594 能够以大幅缩小的尺寸、更低的功耗和整体成本,实现可扩展医疗仪器系统的开发。
支持的产品和硬件
驱动程序或库

AFE159RP4-DESIGN — AFE159RP4 design resources

The full data sheet and other design resources for this device are available here.
支持的产品和硬件
支持软件

TIDA-010280-DESIGN — TIDA-010280 user guide, design files, and firmware

With these downloadable design files, the TIDA-010280 reference design allows users to connect with the TI Simplelink Conncet mobile app for a real-time ECG, Respiration, Pace pulse and temperature demo.
支持的产品和硬件
支持软件

TIDA-010288-DESIGN — TIDA-010288 user guide, design files, and firmware

With these downloadable design files, the TIDA-010288 reference design allows users to connect with the TI Simplelink Conncet mobile app for a real-time AI-enabled ECG, Respiration, Pace pulse and temperature demo.
支持的产品和硬件
参考设计

TIDA-010280 — 具有呼吸、起搏和体温检测功能的小型无线 6 导联 ECG 动态心电图监测仪参考设计

该可穿戴生物传感动态心电图监测仪参考设计提供了一个评估平台,用于评估 TI 持续监测心电图 (ECG)、心率、呼吸、起搏脉冲、体温和运动等生命体征的最新产品。该设计利用 AFE159RP4 通过最多 6 个导联(4 个通道)进行低功耗、高分辨率和高度集成的 ECG 信号采集,并利用 TMP119 进行体温监测。测量数据由 CC2674R10(支持低功耗 Bluetooth® 5.3)传输至智能手机或医疗监测系统等远程终端,以实现实时显示。板载发光二极管 (LED) 可指示系统状态,例如导联脱落、电量低以及低功耗 Bluetooth® 连接。整个系统可以由 2 颗 CR2032 电池(3V (...)

支持的产品和硬件
参考设计

TIDA-010288 — 支持边缘 AI 的无线 ECG 动态心电图监护仪参考设计,适用于实时心律失常分类

该支持边缘人工智能 (AI) 可穿戴生物传感动态心电图监测仪参考设计提供了一个评估平台,用于评估持续监测心电图 (ECG)、心率、呼吸、起搏脉冲、体温和运动等生命体征的最新产品。该设计利用 AFE159RP4 通过最多 6 个导联(4 个通道)进行低功耗、高分辨率和高度集成的 ECG 信号采集,并利用 TMP119 进行体温监测。测量的数据由 MSPM0 MCU 进行处理,借助硬件加速进行实时心律失常分类,然后由 CC2340R5(支持低功耗 Bluetooth® (LE) 5.3)传输到智能手机或医疗监测系统等远程终端以进行实时显示。整个系统可以由 2 节 CR2032 电池(3V (...)
支持的产品和硬件
封装 引脚 CAD 符号、封装和 3D 模型
VQFN (RGZ) 48 Ultra Librarian
DSBGA (YAF) 49 Ultra Librarian

订购和质量

推荐产品可能包含与 TI 此产品相关的参数、评估模块或参考设计。

支持和培训

可获得 TI 工程师技术支持的 TI E2E™ 论坛

所有内容均由 TI 和社区贡献者按“原样”提供,并不构成 TI 规范。请参阅使用条款

如果您对质量、包装或订购 TI 产品有疑问,请参阅 TI 支持。​​​​​​​​​​​​​​

视频