SLVK263 May   2026 TPS7H5030-SP , TPS7H5031-SP

 

  1.   1
  2.   2
  3.   Trademarks
  4. Introduction
  5. Single-Event Effects (SEE)
  6. Device and Test Board Information
  7. Irradiation Facility and Setup
  8. LETEFF and Range Calculation
  9. Test Setup and Procedures
  10. Destructive Single-Event Effects (DSEE)
    1. 7.1 Single-Event Latch-up (SEL) Results
    2. 7.2 Single-Event Burnout (SEB) and Single-Event Gate Rupture (SEGR) Results
  11. Single-Event Transients (SET)
  12. Event Rate Calculations
  13. 10Summary
  14.   A References

Single-Event Transients (SET)

SET are defined as heavy-ion-induced transients upsets on the GATE (OUTH and OUTL tied together) of the TPS7H503x-SP.

Testing was performed at room temperature (no external temperature control applied). The heavy-ions species used for the SET testing was 169Tm at 19.5MeV/nucleon (for more details refer to Table 5-1). Flux of approximately 105ions/cm2/s and a fluence of approximately 107ions/cm2, per run were used for the SET characterization discussed in this chapter.

Waveform size, sample rate, trigger type, value, and signal for all scopes used is presented on Table 8-1.

Table 8-1 Scope Settings

Scope Model

Trigger Signal1

Trigger Type

Trigger Value

Record Length

Sample Rate

PXIe-5110

GATE

Pulse-Width±20%

50k

100MS/s

Open-Loop Configuration

The primary focus of SETs were heavy-ion-induced transient upsets on output signal GATE (OUTH and OUTL tied together). SET testing was done at room temperature at 169Tm which produced a LETEFF of 75MeV×cm2/mg. GATE was monitored using a NI PXIe-5110 and SS was monitored using a NI PXIe-5172. During testing the scope was set to trigger if the signal exceeded |20%| from nominal using a pulse width trigger. During all SET testing, TPS7H503x-SP is SEFI free and there was one type of SET recorded on GATE that was self-recoverable. SS was SET and SEFI free.

The SET results for 4 devices (the same 4 devices used for DSEE) are shown below in the following tables. The transient signature on GATE is shown and the number of transients across the runs, voltages, and frequencies is shown. Since only this transient signature occurred there is high confidence that the TPS7H503x-SP is SEFI free and the recorded transient signature does not show any overshoot.

The upper-bound cross-sections for all bias conditions are shown in Figure 8-1.

Table 8-2 Summary of TPS7H503x-SP Open-Loop SET Test Condition and Results
RUN #UNIT #

Facility

Device Type

PVIN=VIN (V)

FSW (Hz)

IONLETEFF (MeV×cm2/mg)FLUX (ions/cm2/s)FLUENCE (ions/cm2)# GATE ≥ |20%|

13

1

KSEE

TPS7H5030-SP

12

500k

169Tm

75

1.01 × 105

1.00 × 107

44

14

2

KSEE

TPS7H5030-SP

12

500k

169Tm

75

1.01 × 105

1.00 × 107

44

15

3

KSEE

TPS7H5031-SP

12

500k

169Tm

75

9.40 × 104

1.00 × 107

71

16

4

KSEE

TPS7H5031-SP

12

500k

169Tm

75

1.97 × 105

1.00 × 107

28

TPS7H5030-SP Largest Observed GATE Positive Pulse-Width TransientFigure 8-1 Largest Observed GATE Positive Pulse-Width Transient
TPS7H5030-SP Smallest Observed GATE Negative Pulse-Width TransientFigure 8-2 Smallest Observed GATE Negative Pulse-Width Transient
TPS7H5030-SP GATE Pulse-Width Deviation Histogram (All Runs)Figure 8-3 GATE Pulse-Width Deviation Histogram (All Runs)
Table 8-3 TPS7H503x-SP SET Cross-Sections

LETEFF (MeV×cm2/mg)

Frequency (Hz)

VIN

(V)

Fluence (ions/cm2)

# Transients

Upper-Bound Cross-Section (cm2)

75

500k12

4 × 107

187

5.40 × 10-6