AFE4960P

활성

일체형 호흡 및 속도 감지 장치가 포함된 2채널 ECG 및 PPG 아날로그 프런트 엔드

제품 상세 정보

Applications ECG patch, EEG, HRV, Optical-HRM for wearables, PTT-based blood pressure, Pace detection, Patient monitoring, Respiration, Sleep study monitor, SpO2, Vo2 max Number of input channels 6 Resolution (Bits) 22 Number of LEDs 4 Interface type Photodiode Operating temperature range (°C) -40 to 85 Rating Catalog
Applications ECG patch, EEG, HRV, Optical-HRM for wearables, PTT-based blood pressure, Pace detection, Patient monitoring, Respiration, Sleep study monitor, SpO2, Vo2 max Number of input channels 6 Resolution (Bits) 22 Number of LEDs 4 Interface type Photodiode Operating temperature range (°C) -40 to 85 Rating Catalog
DSBGA (YBG) 36 5.76 mm² (0 mm × 0 mm)
  • Supports ECG and respiration impedance measurement; configurable as 2-channel ECG or 1-channel ECG + 1-channel respiration
  • Can be used in IEC 60601-2-47:2012/(R)2016 and IEC 60601-2-27:2011(R)2016 compliant systems
  • 222 µA/channel while operating as 2-channel ECG
  • Integrated Pace pulse detection on 1 channel
  • Supports a 3-lead ECG, extendable to 5 leads or higher by operating two or more AFEs in parallel
  • PPG signal chain for SpO2, Heart rate and PTT
  • ECG signal chain:
    • Single channel ECG acquisition up to 2.048 kHz
    • 2 channel ECG acquisition up to 1.36 kHz/channel
    • RLD output to set body bias through third electrode
    • Programmable INA gain from approximately 2 to 12
    • >1 GΩ input impedance, CMRR > 100 dB
    • Input noise (0.5-150 Hz): 13 µVpp at INA gain of 3; 5 µVpp at INA gain of 12
    • Integrated 370 Hz anti-aliasing low pass filter
    • Mode for continuous lead on/off detection
    • Mode for lead impedance measurement
  • Bio-Z signal chain:
    • Measurement of Bio-impedance over excitation frequency from 30 kHz to 100 kHz
    • Sine wave or square wave excitation
    • Respiration impedance measurement: 45 mΩ-pp noise on a baseline impedance of 2 kΩ
  • Dual channel ECG channel:
    • Bio-Z receiver configurable as a 2nd ECG channel
  • PPG Signal Chain
    • Supports acquisition of up to 24 PPG signals
    • Flexible allocation of 4 LED, 4 PD in each phase
    • 8-Bit LED Current control up to 250 mA
    • Low noise current DAC at input to cancel DC
    • Trans-impedance Gain: 3.7 kΩ to 1 MΩ
    • Filter to reduce sensor’s optical noise bandwidth
  • External clock and internal oscillator modes
  • FIFO with 128-sample Depth, 24-bit word
  • SPITM, I2C interfaces: Selectable by pin
  • 2.6-mm × 2.6-mm DSBGA, 0.4-mm Pitch
  • Supplies: Rx:1.7-1.9 V, IO:1.7-1.9 V , Tx: 3.0-5.5 V
  • Supports ECG and respiration impedance measurement; configurable as 2-channel ECG or 1-channel ECG + 1-channel respiration
  • Can be used in IEC 60601-2-47:2012/(R)2016 and IEC 60601-2-27:2011(R)2016 compliant systems
  • 222 µA/channel while operating as 2-channel ECG
  • Integrated Pace pulse detection on 1 channel
  • Supports a 3-lead ECG, extendable to 5 leads or higher by operating two or more AFEs in parallel
  • PPG signal chain for SpO2, Heart rate and PTT
  • ECG signal chain:
    • Single channel ECG acquisition up to 2.048 kHz
    • 2 channel ECG acquisition up to 1.36 kHz/channel
    • RLD output to set body bias through third electrode
    • Programmable INA gain from approximately 2 to 12
    • >1 GΩ input impedance, CMRR > 100 dB
    • Input noise (0.5-150 Hz): 13 µVpp at INA gain of 3; 5 µVpp at INA gain of 12
    • Integrated 370 Hz anti-aliasing low pass filter
    • Mode for continuous lead on/off detection
    • Mode for lead impedance measurement
  • Bio-Z signal chain:
    • Measurement of Bio-impedance over excitation frequency from 30 kHz to 100 kHz
    • Sine wave or square wave excitation
    • Respiration impedance measurement: 45 mΩ-pp noise on a baseline impedance of 2 kΩ
  • Dual channel ECG channel:
    • Bio-Z receiver configurable as a 2nd ECG channel
  • PPG Signal Chain
    • Supports acquisition of up to 24 PPG signals
    • Flexible allocation of 4 LED, 4 PD in each phase
    • 8-Bit LED Current control up to 250 mA
    • Low noise current DAC at input to cancel DC
    • Trans-impedance Gain: 3.7 kΩ to 1 MΩ
    • Filter to reduce sensor’s optical noise bandwidth
  • External clock and internal oscillator modes
  • FIFO with 128-sample Depth, 24-bit word
  • SPITM, I2C interfaces: Selectable by pin
  • 2.6-mm × 2.6-mm DSBGA, 0.4-mm Pitch
  • Supplies: Rx:1.7-1.9 V, IO:1.7-1.9 V , Tx: 3.0-5.5 V

The AFE4960P supports synchronized signal acquisition of ECG, PPG and bio-impedance signals. A PPG signal chain comprises of a current driver for LEDs and a receiver that senses the signal from PDs. The AFE can be configured as either a 2-channel ECG receiver or as a 1-channel ECG receiver and a Respiration impedance channel. The AFE signal chains can interface in a flexible manner to up to 4 electrodes. The AFE has AC and DC Lead biasing and lead on/off detection. Pacemaker pulse detection is supported on one channel. All the signal chain outputs are converted by a single ADC in well-defined time slots and stored as 24-bit words in a 128-sample FIFO which can be read out using a SPI or an I2C interface. The AFE4960P can be used to realize a 3-lead ECG system along with SpO2.

The AFE4960P supports synchronized signal acquisition of ECG, PPG and bio-impedance signals. A PPG signal chain comprises of a current driver for LEDs and a receiver that senses the signal from PDs. The AFE can be configured as either a 2-channel ECG receiver or as a 1-channel ECG receiver and a Respiration impedance channel. The AFE signal chains can interface in a flexible manner to up to 4 electrodes. The AFE has AC and DC Lead biasing and lead on/off detection. Pacemaker pulse detection is supported on one channel. All the signal chain outputs are converted by a single ADC in well-defined time slots and stored as 24-bit words in a 128-sample FIFO which can be read out using a SPI or an I2C interface. The AFE4960P can be used to realize a 3-lead ECG system along with SpO2.

다운로드 스크립트와 함께 비디오 보기 비디오

추가 정보 요청

전체 데이터시트 및 기타 설계 리소스는 NDA에 따라 제공됩니다. 지금 요청

관심 가지실만한 유사 제품

open-in-new 대안 비교
비교 대상 장치와 동일한 기능을 지원하는 핀 대 핀.
AFE4960 활성 일체형 호흡 및 속도 감지 장치가 포함된 2채널 ECG 아날로그 프런트 엔드 Similar design without integrated photoplethysmography (PPG) analog front end

기술 자료

검색된 결과가 없습니다. 검색어를 지우고 다시 시도하세요.
9개 모두 보기
상위 문서 유형 직함 형식 옵션 최신 영어 버전 다운로드 날짜
* 데이터 시트 AFE4960 Two-Channel ECG, Respiration and Pace Pulse Detection Analog Front End (AFE) for Clinical Wearables datasheet PDF | HTML 2022. 5. 13
Application brief Wearable EEG (Electroencephalography) (Rev. B) PDF | HTML 2023. 10. 30
Product overview ECG and Cuffless Blood Pressure Monitoring With Integrated AFE PDF | HTML 2023. 10. 30
Application brief Remote Patient Monitoring PDF | HTML 2022. 11. 30
Application brief ECG Signal Acquisition on Wearables PDF | HTML 2022. 11. 15
Application brief Measuring Oxygen Saturation (SpO2) on Wearables PDF | HTML 2022. 11. 14
Application brief Optical Heart Rate Monitoring (OHRM) on Wearables PDF | HTML 2022. 11. 14
Application brief Cuffless Blood Pressure Monitoring on Wearables PDF | HTML 2022. 11. 9
인증서 AFE4960PEVM EU RoHS Declaration of Conformity (DoC) 2022. 7. 29

설계 및 개발

추가 조건 또는 필수 리소스는 사용 가능한 경우 아래 제목을 클릭하여 세부 정보 페이지를 확인하세요.

평가 보드

AFE4960PEVM — 호흡 및 속도 감지 장치가 포함된 2채널 ECG 및 PPG AFE용 AFE4960 평가 모듈

AFE4960P 평가 모듈(EVM)은 AFE4960P 바이오센싱 AFE(아날로그 프런트 엔드) 평가를 위한 플랫폼으로, 심전도(ECG), 생체 임피던스 및 광전용적맥파 측정 신호의 동기화된 신호 수집을 지원합니다.

AFE4960은 PPG 채널과 함께 2채널 ECG 수신기 또는 1채널 ECG 수신기 및 호흡 임피던스 채널로 구성할 수 있습니다. AFE 신호 체인은 최대 4개의 전극까지 유연한 방식으로 상호 작용할 수 있습니다. RLD(Right-leg Drive) 증폭기 출력을 사용하여 바디 바이어스를 설정할 수 있습니다. (...)

지원되는 제품 및 하드웨어
시뮬레이션 툴

PSPICE-FOR-TI — TI 설계 및 시뮬레이션 툴용 PSpice®

TI용 PSpice®는 아날로그 회로의 기능을 평가하는 데 사용되는 설계 및 시뮬레이션 환경입니다. 완전한 기능을 갖춘 이 설계 및 시뮬레이션 제품군은 Cadence®의 아날로그 분석 엔진을 사용합니다. 무료로 제공되는 TI용 PSpice에는 아날로그 및 전력 포트폴리오뿐 아니라 아날로그 행동 모델에 이르기까지 업계에서 가장 방대한 모델 라이브러리 중 하나가 포함되어 있습니다.

TI 설계 및 시뮬레이션 환경용 PSpice는 기본 제공 라이브러리를 이용해 복잡한 혼합 신호 설계를 시뮬레이션할 수 있습니다. 레이아웃 및 제작에 착수하기 (...)
지원되는 제품 및 하드웨어
패키지 CAD 기호, 풋프린트 및 3D 모델
DSBGA (YBG) 36 Ultra Librarian