AFE4960
- Supports ECG and respiration impedance measurement; configurable as 2-channel ECG or 1-channel ECG + 1-channel Respiration
- Can be used in IEC 60601-2-47:2012/(R)2016 and IEC 60601-2-27:2011(R)2016 compliant systems
- 222 µA/channel while operating as 2-channel ECG
- Integrated Pace pulse detection on 1 channel
- Supports a 3-lead ECG, extendable to 5 leads or higher by operating two or more AFEs in parallel
- ECG signal chain:
- Single channel ECG acquisition up to 2.048 kHz
- 2 channel ECG acquisition up to 1.36 kHz/channel
- RLD output to set body bias through third electrode
- Programmable INA gain from 2~12
- >1 GΩ input impedance, CMRR > 100 dB
- Input noise (0.5-150 Hz): 13 µVpp at INA gain of 3; 5 µVpp at INA gain of 12
- Integrated 370 Hz anti-aliasing low pass filter
- Mode for Continuous lead on/off detection
- Mode for Lead impedance measurement
- Bio-Z signal chain:
- Measurement of Bio-impedance over excitation frequency from 30 kHz to 100 kHz
- Sine wave or square wave excitation
- Respiration impedance measurement: 45 mΩ-pp noise on a baseline impedance of 2 kΩ
- Dual channel ECG channel:
- Bio-Z receiver configurable as a 2nd ECG channel
- External clock and internal oscillator modes
- FIFO with 128-sample Depth, 24-bit word
- SPITM, I2C interfaces: Selectable by pin
- 2.6-mm × 2.6-mm DSBGA, 0.4-mm Pitch
- Supplies: Rx:1.7-1.9 V, IO:1.7-1.9 V
The AFE4960 can be configured as either a 2-channel ECG receiver or as a 1-channel ECG receiver and a respiration impedance channel. The AFE signal chains can interface in a flexible manner to up to 4 electrodes. A Right Leg Drive (RLD) amplifier output, can be used to set the body bias. The AFE has DC lead biasing for lead on/off detection, and AC lead biasing for measuring the lead impedance. Pacemaker pulse detection is supported on one channel.
All the signal chain outputs are converted by a single ADC in well-defined time slots and stored as 24-bit words in a 128-sample FIFO which can be read out using a SPI or an I2C interface.
The AFE4960 is a fully integrated solution to realize a 3-lead ECG system. Synchronized operation of two AFEs in parallel can be used to realize a 5-lead ECG.
추가 정보 요청
전체 데이터시트 및 기타 설계 리소스는 NDA에 따라 제공됩니다. 지금 요청
기술 자료
| 상위 문서 | 유형 | 직함 | 형식 옵션 | 최신 영어 버전 다운로드 | 날짜 |
|---|---|---|---|---|---|
| * | 데이터 시트 | AFE4960 Two-Channel ECG, Respiration and Pace Pulse Detection Analog Front End (AFE) for Clinical Wearables datasheet (Rev. A) | PDF | HTML | 2022. 3. 7 | |
| Application brief | Wearable EEG (Electroencephalography) (Rev. B) | PDF | HTML | 2023. 10. 30 | ||
| Application brief | Remote Patient Monitoring | PDF | HTML | 2022. 11. 30 | ||
| Application brief | ECG Signal Acquisition on Wearables | PDF | HTML | 2022. 11. 15 | ||
| Application brief | Analog Front End for 3-Lead and 5-Lead ECG (Rev. A) | PDF | HTML | 2022. 4. 20 |
설계 및 개발
추가 조건 또는 필수 리소스는 사용 가능한 경우 아래 제목을 클릭하여 세부 정보 페이지를 확인하세요.
AFE4960EVM — 일체형 호흡 및 속도 감지 장치가 포함된 2채널 ECG AFE용 AFE4960 평가 모듈
AFE4960 평가 모듈(EVM)은 AFE4960 바이오센싱 AFE(아날로그 프론트 엔드)를 평가하기 위한 플랫폼으로, 심전도(ECG) 및 생체 임피던스 신호의 동기화된 신호 수집을 지원합니다.
AFE4960은 2채널 ECG 수신기 또는 1채널 ECG 수신기 및 호흡 임피던스 채널로 구성할 수 있습니다. AFE 신호 체인은 최대 4개의 전극까지 유연한 방식으로 상호 작용할 수 있습니다. RLD(Right-leg Drive) 증폭기 출력을 사용하여 바디 바이어스를 설정할 수 있습니다. AFE에는 리드 온/오프 감지를 위한 DC 리드 (...)
TIDA-010270 — Bluetooth® 저에너지를 사용한 지속적인 생명 징후 모니터링을 위한 바이오센싱 패치 레퍼런스 설계
바이오센싱 패치 레퍼런스 설계는 심전도(ECG), 심박수, 호흡, 페이스 펄스, 온도 및 동작과 같은 생명 징후를 지속적으로 모니터링하기 위한 최신 제품을 평가할 수 있는 플랫폼을 제공합니다. 이 설계는 정확한 단일 리드 ECG 신호 수집에 AFE4960을, 및 체온 모니터링에 TMP119를 사용합니다. 측정된 데이터는 CC2674R10에서 스마트폰 및 의료 모니터링 시스템과 같은 원격 터미널로 전송되어 실시간 디스플레이를 지원합니다. 온보드 LED(발광 다이오드)는 리드 오프, 저전력 및 Bluetooth® LE(저전력) 연결과 (...)
| 패키지 | 핀 | CAD 기호, 풋프린트 및 3D 모델 |
|---|---|---|
| DSBGA (YBG) | 36 | Ultra Librarian |
주문 및 품질
권장 제품에는 본 TI 제품과 관련된 매개 변수, 평가 모듈 또는 레퍼런스 디자인이 있을 수 있습니다.